21
TissueWave 2 Operator Guide
Air agitation during processing helps in eliminating cold and hot
spots.
The constant air flow causes the air to circulate and prevents hot and
cold spots from affecting any one particular area.
The total
Air Agitator
flow rate is
1 l/min Nominal
.
The
Air Agitator Pump
may be controlled by the
Air Agitator Key
located on the
Keypad
.
The
Pump
is usually controlled by a command in the
Protocol Edit
screen.
However, the operator may override this setting with the
Air Agitator
Key
.
The
Air Agitator Outlet Tube
may be removed from the bushing and
placed separately in the container for agitation.
The volume of the air stream producing the agitation is controlled
manually with the
Flow Control Valve Knob
located at the
connection of the tubing to the unit.
The airflow is
Decreased
as the knob is turned C
lockwise
.
The
Interior Light
can be turned on and off using the
Interior Light
Key
on the
Keypad
.
This function can be used at any time while the instrument is
switched
On
at the Mains.
Air
Agitator
Air
Agitator
Interior
Light
Interior
Light