19
TissueWave 2 Operator Guide
The Shandon TissueWave 2
Vacuum Accumulator
is designed to
prevent the accidental ingestion of liquid paraffin or other
contaminants by the system’s
Vacuum Pump
.
Should liquids enter the
Vacuum System
via the
Vacuum Inlet
Connection
inside the
Processor Cavity
, they will be discharged
into the
Accumulator Canister
and collected at the bottom.
Paraffin will solidify and remain there until re-heated for removal.
The accumulator’s collection capacity before vacuum operation is
affected is approximately
15 fl oz (450 ml)
.
The flow rate of the
Vacuum Pump
is
2200 in³/min (0.036 m³/min)
.
The
Flexible Temperature Probe
may be used in conjunction with
the
Air Agitator
or threaded through the
Vacuum Chamber
during
Paraffin Infiltration
.
Vacuum
Accumulator
Vacuum
Accumulator
Temperature
Probe
Temperature
Probe