There are two safety interlocks on the door to prevent open-
door operation.
Do not allow residue to build up on the door seals. Seals may
be cleaned with a mild detergent and warm water using a
soft sponge or cloth.
Do not operate the unit if the door, hinges or door seals are
damaged.
Use caution when heating volatile and / or flammable
materials.
Volatile material may evaporate quickly in the microwave
chamber and should be closely monitored during processing.
To prevent the build-up of an explosive mixture in the
chamber, flammable materials should not be heated to their
boiling point.
Reagents containing metals (eg: Zinc Formalin ot Mercuric
Chloride) need to be completely cleaned off of all internal
components. Build up of such metals may cause arcing to
occur.
Do not install the unit next to or near a source of heat.
Do not install the unit at or above eye level. Make sure the
microwave fits securely on a counter space.
To avoid toxic fumes when handling heated solutions always
work with heated solutions in a properly vented area.
Never use an additional thermometer in the unit.
Only the supplied temperature probe should be used to
monitor temperature inside of the Shandon TissueWave 2.
Use only qualified microwave service personnel for
repairs.
•
•
•
•
•
•
•
•
•
•
!!
!!
!!