Fault Diagnosis
Alarm Conditions
9-2
Prima PRO & Sentinel PRO Mass Spectrometers User Guide
Thermo Fisher Scientific
Device
Typical
Reading
Set Point
Low
Set Point
High
Typical Cause
ASU Controller Status
1
0.5
1.5
ASU failure
Cabinet Temperature
24.00°C
10.00°C
35.00°C
AC fault or AC fan failed /
blocked
Electronics
Temperature
28.20°C
10.00°C
55.00°C
Ambient temperature or
cabinet filter blocked
Magnet Controller
Status
1
0.5
7
Magnet busy or failure
Mass Filter
0.45690
Tesla
Pressure Set Point
3.44e-05
mbar
System Pressure
3.13e-06
mbar
2.00e-07
mbar
2.50e-05
mbar
Still pumping or leak or
pump failure
System Pressure Trip
(OK)
(OK)
Still pumping or leak or
pump failure
Turbo Disable
(Off)
(Off)
Turbo switched to “stop”
on PDC
Turbo Motor Current
0.62 amps
0.01 amps
4.00 amps
Turbo adjustment or
failure
Turbo Operating
Hours
7264 hours
Turbo Speed
100%
95%
Still pumping or leak or
pump failure
Turbo Speed Interlock
(OK)
Still pumping or leak or
pump failure
Vacuum
(OK)
(OK)
Still pumping or leak or
pump failure
Vent Valve Off
(Off)
RMS Body Temp
59°C
30°C
100°C
Still heating up or heater
failure
RMS Ambient
Temperature
42°C
RMS Controller Status
1
0.5
1.5
RMS electronics failure
RMS Digital Input 2
(Fail)
RMS Digital Input 3
(Fail)
RMS H1 Current
245.0 mA
50.0 mA
Heater failure
RMS Sample Tube
Current
350.0 mA
50.0 mA
Heater failure
RMS Flow Sensor
(OK)
(OK)
Low sample flow
RMS Position
31
0.5
Optical sensor / detector
needs cleaning