Status & Tuning Parameters
Communications
Thermo Fisher Scientific
Prima PRO & Sentinel PRO Mass Spectrometers User Guide
8-11
Current passing through the electromagnet coils can be used to give
an approximate indication of the current mass number. The
approximate current to mass relationship is 4 [I
2
] for systems with
1000 V ion energy.
Mass 100 therefore requires roughly 5 A through the coils. This
relationship is very approximate. If the mass table has not yet been
generated (or has been corrupted), this relationship can be used to
check whether or not the magnet controls circuitry is operating.
The Magnet Sync AC parameter is an AC power synchronization
signal to the magnet driver card, which improves precision and
reduces mains frequency induced noise.
The Power Controller Status parameter indicates a value of
1
when
unit is operating within normal limits.
The Turbo 24V parameter is the 24 V PDU output for the turbo
molecular pump drive electronics.
The User IO 24V parameter is the 24 V PDU output for the user I/O.
The Vent Valve 24V parameter is the 24 V PDU power supply for
the turbo molecular pump vent valve.
The Host Comms Quiet parameter checks for a break in
communications between the host PC and the Prima PRO analyser.
If configured (within GasWorks Hardware Configuration), the User
I/O 0 Comms Quiet status is checking for a break in communications
between the user I/O card port 0 and a user device.
If configured (within GasWorks Hardware Configuration), the User
I/O 1 Comms Quiet status is checking for a break in communications
between the user I/O card I/O 1 port and a user device.
If configured (within GasWorks Hardware Configuration), the User
I/O 2 Comms Quiet status is checking for a break in communications
between the user I/O card I/O 2 port and a user device.
Magnet Supply
Current
Magnet Sync AC
Power Controller
Status
Turbo 24V
User IO 24V
Vent Valve 24V
Communications
Host Comms
Quiet
User I/O 0
Comms Quiet
User I/O 1
Comms Quiet
User I/O 2
Comms Quiet