Maintenance
Procedures
10-2
Prima PRO & Sentinel PRO Mass Spectrometers User Guide
Thermo Fisher Scientific
Warning! High Temperatures!
Rotary pumps run hot during operation.
Allow pumps to cool before draining oil.
The oil level in the rotary pump should be checked in accordance
with the maintenance schedule and replenished if necessary. If a
significant loss of oil has occurred, this should be investigated
according to instructions provided by the pump manufacturer
(manual supplied).
Routine oil changes are not required for rotary pumps filled with
Fomblin oil.
Figure 10
–1.
Rotary pump drain plug
To maintain adequate cooling of the instrument, ensure that the fan
covers, air intake, and condenser coil are free of dust. Clean as
required.
Figure 10
–2.
Fan cover guard
Cleaning the
Cooling Fans
Fan cover guard
Drain plug