Chapter 7
Startup & Shutdown
These procedures assume that the system has been installed and
commissioned as earlier in this manual.
The startup procedure for the instrument is as follows:
1.
Ensure that the rotary pump is filled with oil to a point just below
the upper marker on the oil level window. Ensure that the pump
outlet is not obstructed.
2.
Check that the Main Electrical ISOLATOR is switched OFF:
For safe area systems, red / yellow rotary switch on lower left
side of cabinet, OFF = HORIZONTAL.
For hazardous area systems, red E-Stop switch on left side of top
mounted flameproof enclosure pressed in.
All the internal circuit breakers should be switched to OFF
(breaker to down position).
Check that the TURBO RUN toggle switch (middle of the front
of the PDU) is in the OFF position (OFF = switch UP!).
3.
Apply mains power to the instrument and switch the Mains
Electrical ISOLATOR to ON:
For safe area systems, red rotary switch ON = VERTICAL.
For hazardous area systems, E-Stop switch ON = OUT (rotate to
reset, key required).
In the case of an Ex system, an initial purge cycle will need to be
completed (automatic or manual, depending on area
classification) before mains power is switched to the instrument
cabinet.
4.
Switch the circuit breakers to ON (breaker to UP position):
CABINET, INLET, ROTARY 1 (and ROTARY 2, if applicable),
and COOLING.
5.
Turn the TURBO RUN toggle switch to ON (ON = switch
DOWN!).
Startup
Procedure