Safety Information
System Safety
Thermo Fisher Scientific
Prima PRO & Sentinel PRO Mass Spectrometers User Guide
2-5
Warning!
The following assemblies may attain a surface
temperature as listed:
Rapid Multi-Stream Sampler (RMS) and Single Point Inlet
System: 120°C.
Single Point Inlet System (High Temperature Version):
230°C.
RMS Sample Probe: 120°C.
Inlet Probe Assembly: 120°C.
Rotary pump operating temperature: 75°C.
Warning!
Ensure that the rotary pump is unobstructed. Failure to
do so
will
cause dangerous internal pressures in the pump. This
will
result in blown seals and bursting of the oil box and could result in
physical injury.
Warning!
A leak-tight exhaust must be fitted to the inlet system
when sampling gas streams which are:
Explosive
Toxic
Asphyxiant
High
Temperatures
Rotary Pump
Exhaust Outlet
Inlet System
Exhaust