PerpHecT
®
Orion 310, 320, 330 Instruction Manual
44 General Information on pH Calibration for Orion 330
Calibration and Measurement of pH
A. Autocalibration with Two Buffers
1. Attach the pH electrode to the meter.
2. Choose either pH 4.01 and pH 7.00, or pH 7.00 and pH 10.01
buffers; whichever will bracket your expected sample range.
3. Press CAL key to initiate calibration sequence. CAL is displayed
for two seconds. Press the YES key to accept the last calibration
range (7-4 or 7-10) or select one of these calibration ranges using
the SCROLL (
▲
,
▼
) keys. Press the YES key to accept. The 7
buffer annunciator will light up. Place electrode in pH 7.00 buffer.
Reading will be displayed and updated as calibration continues.
When the READY light comes on, indicating electrode stability,
press the YES key to accept.
4. The pH 4 (or pH 10) annunciator will light up. Remove electrode
from pH 7.00 buffer. Rinse with deionized water and place
electrode in either pH 4.01 or 10.01 buffer (depending on
calibration range that was selected). When the READY light
comes on, press the YES key to accept buffer value. SLP will be
displayed while the calculated slope is displayed. If a printer is
attached, calibration data will be sent to the printer.
5. The annunciators for the type of calibration performed will remain
lit until another calibration is performed.
6. Meter will automatically go into the MEASURE mode.
The READY light will come on to indicate electrode stability.
7. If operating in the LogR mode or using an ATC probe, the
temperature-corrected pH reading is displayed.
8. Rinse electrodes with deionized water and place into sample.
Record pH directly from the main meter display when the READY
light is displayed or when electrode signal is stable.
NOTE:
When making measurements in LogR
mode, you must calibrate for temperature before
attempting a pH calibration. See page 49 for
temperature calibration procedure.
Содержание Orion 310
Страница 12: ...PerpHecT Orion 310 320 330 Instruction Manual 8 ...
Страница 18: ...PerpHecT Orion 310 320 330 Instruction Manual 14 ...
Страница 24: ...PerpHecT Orion 310 320 330 Instruction Manual 20 ...
Страница 26: ...PerpHecT Orion 310 320 330 Instruction Manual 22 ...
Страница 30: ...PerpHecT Orion 310 320 330 Instruction Manual 26 ...
Страница 40: ...PerpHecT Orion 310 320 330 Instruction Manual 36 ...
Страница 42: ...PerpHecT Orion 310 320 330 Instruction Manual 38 ...
Страница 46: ...PerpHecT Orion 310 320 330 Instruction Manual 42 ...
Страница 56: ...PerpHecT Orion 310 320 330 Instruction Manual 52 ...
Страница 64: ...PerpHecT Orion 310 320 330 Instruction Manual 60 Troubleshooting Guide ...
Страница 66: ...PerpHecT Orion 310 320 330 Instruction Manual 62 ...
Страница 76: ...PerpHecT Orion 310 320 330 Instruction Manual 72 ...
Страница 78: ...PerpHecT Orion 310 320 330 Instruction Manual 74 ...