Description of the device
4.7
Sample chamber illumination
Fig. 8:
The sample chamber illumination unit [2]:
•
one flourescent tube for model MSC 1.2,
•
two flourescent tubes for model MSC 0.9, 1.5,1.8,
is installed behind the light dome [3].
4.8
UV lamp unit
Fig. 8:
The optional, device-integral UV lamp [1] is installed
to the ceiling at the front section of the sample chamber.
The operating time of the UV lamp is preset. The UV disinfection
routine can be started by depressing a key on the operating
panel.
NOTE – Protection from UV radiation!
To protect from UV radiation, the UV lamp can
be activated only when the front window is
completely closed.
When the front window is opened, the UV di-
sinfection routine is cancelled immediately.
4.9
Working area
The standard equipment comprises the segmented work-
plate. The workplate segments are placed onto the frame above
the sample chamber floorpan.
NOTE – Workplate condition
Use properly plane lying workplates only.
Replace deformed workplates. Pay attention
to the maximum work load (see sction 12)!.
Fig. 9:
The working area
A
for optimal product protection extends
over the entire width
B
and depth
C
of the workplate [1]. The two
armrests [3] are positioned centrically to the working area at
distance
D
(20 cm) from each other. For attachment, each arm-
rest is inserted into the first perforation track [2] of the workplate.
NOTE – Using the armrests
Safe working in the work mode of the device
(see section 6.3) is ensured only if the armrests
are used correctly!
Fig.9
Working area on the workplate,
armrests
Fig.8
UV lamp unit