15
Operating Instructions Biological Safety Cabinet | MSC-Advantage
3
Installation
3.5
Transport
Fig. 2:
To prevent tilting, always transport the device using a sui-
table carrier, even for a transport within a building, and separate
it from the stand (see Section 5.2).
CAUTION – Tilting danger!
If the unit is tilted too much during lifting,
risk of tipping exists. Lift the biological safety
cabinet only
vertically!
For transportation (including inside buildings)
use an appropriate lifting device, which
ensures that the unit:
• is on a stable stand and
• is secured against lateral tilting.
Do not transport the biological safety cabinet
on a base with roles.
CAUTION – Lift points!
For transport, lift the device only at the lift
points shown in Figure 2.
Do not allow the weight of the cabinet to rest
on the floorpan!
CAUTION – Crushing hazard!
The weight of the front window is counterba-
lanced by the counterweight (at the rear of the
device).
The device must not be transported unless the
counterweight has been locked.
•
Install the four lockscrews (see Section 5.3).
Fig. 2
Lift points