Shut-down
8.1
Interrupting an operation
To interrupt a work process:
1. Remove all samples from the biological safety cabinet and store them properly.
2. Remove accessories from the sample chamber and clean and disinfect them.
3. Clean and disinfect the sample chamber surfaces, the workplate, and the
floorpan.
4. Switch the device to standby mode: Close the front window and keep the ON
key depressed until the indicators are off (the right indicator segment of the
display shows a dot to indicate that voltage is present).
NOTE – Blower switch-off!
For safety reasons, the blowers can be swit-
ched off only when the front window is closed.
8.2
Shutting the device down
If the device is not to be used or stored for an extended period of time, it must
be completely decontaminated.
WARNING – Decontamination measures!
To shut the device down, the sample chamber
must be disinfected completely and the ple-
num, including the filters, must be sterilized
using formaldehyde.
1. After the device has been decontaminated, close the front window completely.
2. Disconnect the device from the power supply system.