Handling and control
Standby mode:
The front window can be lowered to seal the sample chamber.
The air system output is reduced to match the lower air requirement.
•
The front window is closed:
– The air system operates at reduced output.
– The red status indicator
FRONT WINDOW NOT IN WORK POSITON
is
illuminated.
– The yellow status indicator
AIRFLOW REDUCED
is illuminated.
– The red status indicator
A
IRFLOW NOT STEADY
is illuminated.
•
The sample chamber illumination is enabled.
•
The internal power supply in the sample chamber is enabled:
– If the internal power supply is ON, the yellow status indicator
INTERNAL
POWER SUPPLY ACTIVATED
is illuminated.
•
The biological safety cabinet can be switched off (in OFF mode):
– Keep the ON key depressed until all indicators are off.
UV mode:
For running the UV disinfection routine, the front window is completely
lowered to protect against UV radiation. The routine cannot be run until the front
window is in the "closed" position.
•
The UV disinfection routine is activated:
– The yellow status indicator
UV DISINFECTION ROUTINE ACTIVATED
is illu-
minated until the preset time for the routine has elapsed. Then, the UV
lamps are switched off automatically, and the status indicator is swit-
ched off.
•
The sample chamber illumination is disabled.
•
The internal power supply in the sample chamber is disabled.
•
The integral UV lamp (optional) is enabled.
•
The power supply for the UV disinfection adapter (optional) is enabled.