45
Operating Instructions Biological Safety Cabinet | MSC-Advantage
10
Maintenance
10.4
Exchanging SmartPorts
NOTE – Contamination
In order to prevent contaminated particles
escaping from the utility chamber, the unit
must be switched on in order to exchange the
SmartPorts.
The SmartPorts are removed inwards into the
utility chamber.
1. First remove the inner grommet of the SmartPort in the utility
chamber inwards from the side panel opening.
2. Then remove the outer grommet of the SmartPort in the utility
chamber inwards from the side panel opening.
3. Disinfect both grommets in the utility chamber and take them
out of the utility chamber.
4. Insert new outer grommet in the side panel opening from the
outside.
5. Insert new inner grommet in the side panel opening from the
inside.