36
Qubit
™
Flex Fluorometer User Guide
Read samples
Before you begin
•
Calibrate the Qubit
™
Flex Fluorometer as described on page 25. (Run the
appropriate standards or accept the values from the previous calibration.)
•
Prepare the samples. Refer to the instructions provided with the assay.
Note:
Incubate the samples for the appropriate amount of time after
mixing them with the working solution (2 minutes for the Qubit
™
DNA
and RNA assays, 15 minutes for the Qubit
™
protein assay).
Insert samples
1.
When prompted, load the tube strip containing the samples as shown in the
Insert samples
screen. If you have fewer than 8 samples, press to deselect the
tube positions that do not contain a sample.
All 8 tubes contain samples
No sample in positions S7 and S8
2.
Press
Output sample units
to open the
Output Units
screen, then select the desired units.