Appendix A: Thermo Scientific iEMS Incubator / Shaker HT Quick Reference Guide
Thermo Scientific iEMS Incubator / Shaker HT User Manual Rev. 2.4, Cat. no. 1506400A
54
Appendix A:
Thermo Scientific iEMS Incubator / Shaker HT Quick
Reference Guide
1.
Place the microplate into the thermal microplate holder so that well A1 is
located in the upper left corner when viewed with the numbers on the
holder's handle the right side up. Refer to Fig. 6.2.
2.
Insert the thermal microplate holder into an incubation slot. Refer to Fig. 6.3
and Fig. 6.4.
3.
Switch the instrument ON. Refer to Fig. 4.1, item 7.
4.
Program the incubation and shaking parameters for the appropriate slot:
Press the slot number (
1
−
3
) and
↵
(Enter). For all slots, press
0
↵
.
Enter the incubation and shaking parameters in the following order (refer
to Section 6.2), and confirm each selection by pressing
↵
:
Temperature (°C)
Incubation time (hh:mm:ss)
Shaking time (hh:mm:ss)
Interval time (hh:mm:ss)
Shaking speed as shown below:
Value
0 1 2 3 4 5
Shake speed (rpm)
0 400 650 900 1150 1400
5.
Press
temp
followed by the slot number to start the preheating of the
thermal microplate holder. Refer to Section 6.5.
6.
Wait until the control LED changes from a flashing orange to a continuous
orange.
7.
Press the slot number (
1
−
3
) and then
start
to start the protocol. The
protocol will start and the LED will light up with a red color.
8.
When the protocol is completed, the LED will light up with a green color
and you will hear a beeping sound. The heating stays on.
9.
To turn off the heating, press the slot number and then
stop
.
10.
Remove the thermal microplate holder from the slot when the module is not
shaking.
Warning:
The thermal microplate holder will be hot after use. Only move the
microplate holder from its handle. Use protective gloves when opening
the cover and removing the microplate.
Caution:
Do not place the microplate holder in or remove it from the instrument
while the instrument is shaking. This may damage the shaking
mechanism.
11.
Turn the instrument OFF.
Содержание iEMS Incubator
Страница 2: ...Thermo Scientific iEMS Incubator Shaker HT User Manual Rev 2 4...
Страница 3: ......
Страница 4: ...Thermo Scientific iEMS Incubator Shaker HT User Manual Rev 2 4 Cat no 1506400A...
Страница 61: ...Thermo Scientific iEMS Incubator Shaker HT User Manual Rev 2 4 Cat no 1506400A 58...
Страница 62: ...Thermo Scientific iEMS Incubator Shaker HT User Manual Rev 2 4 Cat no 1506400A 59...
Страница 63: ...Thermo Scientific iEMS Incubator Shaker HT User Manual Rev 2 4 Cat no 1506400A 60...
Страница 64: ......