Chapter 5: Installation
Thermo Scientific iEMS Incubator / Shaker HT User Manual Rev. 2.4, Cat. no. 1506400A
16
5.2.6
Technical prerequisites
Place the instrument on a normal laboratory bench. The net weight of the
instrument is 15 kg
[
33 lbs.
]
.
The instrument operates at voltages of 100 – 120 Vac, 220 – 240 Vac and the
frequency range 50/60 Hz.
5.3
Setups before you put the instrument into operation
5.3.1
Voltage selector
The voltage selector/fuse block assembly (Fig. 5.1) is located on the left-hand
side (Fig. 4.1, item 8) of the instrument. Check that the selected voltage
corresponds to the local voltage. You can see the current voltage setting
through the slit in the fuse holder (Fig. 5.2, item 3).
To change the selected voltage:
1.
Remove the power cable from the power inlet.
2.
Use a small flat screwdriver to pry the voltage selector/fuse block assembly
away from the power entry port housing and remove (Fig. 5.1).
Fig. 5.1 Opening the voltage selector/fuse block assembly