Chapter 6: Operation
Thermo Scientific iEMS Incubator / Shaker HT User Manual Rev. 2.4, Cat. no. 1506400A
28
Within a module, a second protocol
with a programmed temperature
different from the protocol already in
progress will not start and the display
will show:
When the protocol is completed, the LED will light up with a green color and
you will hear a beeping sound. The heating stays on. For turning off the beeping
sound, see Section 6.4.3.6.
6.4.3.4
Checking the operation stop time
Before taking the thermal microplate holder out of the slot, you can check the
actual time of day when the incubation stopped. Press the slot number (
1
−
3
).
For example,
Press
1
Press
C
to return to the READY state.
If you take the thermal microplate holder out of the slot, you can no longer
check the time indicating the stop of the incubation.
6.4.3.5
Stopping the operation
If you press a slot number and
stop
before the processing protocol time
elapses, the protocol stops and incubation stays on. The LED will light up with a
green color and you will hear a beeping sound. For turning off the beeping
sound, see Section 6.4.3.6.
The READY state appears.
6.4.3.6
Turning off the audio signal
Caution:
Do not place the microplate holder in or remove it from the instrument
while the instrument is shaking. This may damage the shaking
mechanism.
If you want to turn off the audio signal but wish to continue the incubation,
proceed as follows:
1.
Make sure the control LED is green and that the module is not shaking.
2.
Remove the thermal microplate holder from the slot. The audio signal and
the control LED are turned off and the heating of the slot is turned off.
3.
Re-insert the microplate holder into the slot. The orange LED is turned on
and the heating of the slot turns back on.
PLACE 3
PROHIBITED OPERATION
PLACE 1
INC STOP
09:20:31
READY
09:27:30
Содержание iEMS Incubator
Страница 2: ...Thermo Scientific iEMS Incubator Shaker HT User Manual Rev 2 4...
Страница 3: ......
Страница 4: ...Thermo Scientific iEMS Incubator Shaker HT User Manual Rev 2 4 Cat no 1506400A...
Страница 61: ...Thermo Scientific iEMS Incubator Shaker HT User Manual Rev 2 4 Cat no 1506400A 58...
Страница 62: ...Thermo Scientific iEMS Incubator Shaker HT User Manual Rev 2 4 Cat no 1506400A 59...
Страница 63: ...Thermo Scientific iEMS Incubator Shaker HT User Manual Rev 2 4 Cat no 1506400A 60...
Страница 64: ......