Chapter 6: Operation
Thermo Scientific iEMS Incubator / Shaker HT User Manual Rev. 2.4, Cat. no. 1506400A
30
Fig. 6.5 Removing the microplate holder – lifting the handle
Fig. 6.6 Removing the microplate holder – pulling the holder out
3.
Place the microplate holder on a heat resistant surface. Open the cover and
remove the microplate.
Warning:
The thermal microplate holders are designed to stop heating if
temperatures reach 80°C ± 5°C in the unlikely event that temperature
control to a holder is lost or enters a runaway condition. Microplate
holders at this temperature will be hot when touched. To help prevent
loss of temperature control, ensure the power contacts on the back of
the microplate holder are clean and free from dirt or corrosion and that
the holder is correctly inserted into the slot before you start the
processing protocol.
Instrument reset due to power loss
If the power to the instrument is lost, the instrument will reset to the READY
mode when power is restored. All shaking and incubation functions, including
timing, will be turned off. No LEDs will be lit and the current time of day will be
shown on the display.
Содержание iEMS Incubator
Страница 2: ...Thermo Scientific iEMS Incubator Shaker HT User Manual Rev 2 4...
Страница 3: ......
Страница 4: ...Thermo Scientific iEMS Incubator Shaker HT User Manual Rev 2 4 Cat no 1506400A...
Страница 61: ...Thermo Scientific iEMS Incubator Shaker HT User Manual Rev 2 4 Cat no 1506400A 58...
Страница 62: ...Thermo Scientific iEMS Incubator Shaker HT User Manual Rev 2 4 Cat no 1506400A 59...
Страница 63: ...Thermo Scientific iEMS Incubator Shaker HT User Manual Rev 2 4 Cat no 1506400A 60...
Страница 64: ......