Replace the instrument fuses
Materials required
• Fuses, 1.6A, Time-Lag T, 250VAC, 5 × 20-mm (2)
• Safety glasses
• Powder-free gloves
• Screwdriver, flathead
Replace the fuses
CAUTION!
FIRE HAZARD
. For continued protection against the risk of fire, replace
fuses only with listed and certified fuses of the same type and rating as those
currently in the iBright
™
imager.
1.
Power off, then unplug the iBright
™
imager.
2.
Using a flat-head screwdriver, pry open the fuse door, and remove the fuse
holder.
3.
Remove each fuse from its fuse holder and inspect it for damage. Carbon
typically coats the inside of failed fuses.
4.
Replace each failed fuse with a 1.6A, Time-Lag T, 250VAC, 5 × 20-mm Fuse.
Note:
The voltage and amperage ratings are on the fuse holder.
Maintain the instrument
Replace the instrument fuses
iBright
™
CL750 Imaging System User Guide
31