Thermo Scientific HM525 NX Operator Guide
29
Fine Adjustment of the Anti-Roll Device
•
The fine adjustment of the anti-roll device is carried out via the knurling screw.
Selecting the Clearance Angle
•
Loosen the clamping lever.
•
Swivel the upper part of the blade carrier on the base until the desired clearance angle is reached.
•
The clearance angle can be read on the scale on the left side of the base.
•
Bring the clamping lever into clamping position.
•
The selected clearance angle is now fixed in its position.
Note
Usable cuts are only achieved at a clearance angle of 10° or more.
Specimen Head HM525 NX
1
Microtome base plate
2
Electrical connector
3
Specimen Chuck Release Lever
4
Specimen Head Clamping Lever
5
Blade/Knife holder Clamping Lever