Thermo Scientific HM525 NX Operator Guide
36
Set the Specimen Head Position
•
Move the Specimen Head to the desired position and press the “Position” button on the Main
screen for about 3 seconds.
•
The colour will change to yellow, showing that the position has been saved and the required
position is stored.
•
When a new block is inserted into the Specimen Head, press the “Position” button for 1 second
and the Specimen Head will return to the set position.
•
To save a new position, again align the Specimen Head to the desired position and press the
“Position” button for 3 seconds. The new position will be stored.
•
To reset the position, hold down the position button for approx 6 seconds.
Backward fast movement of specimen head (1mm/sec) button
Press the button for 1 sec to move the Specimen Head backwards a short distance.
After about 2 sec the head moves back automatically.
Press the button again to stop the movement immediately.
Slow forward movement of specimen head (0.5mm/sec) button
Press the button to move the Specimen Head forward.
The head will only move while the button is pressed.
It will stop immediately if not pressed.
Fast forward movement of specimen head (1mm/sec) button
Press the button to move the Specimen Head fast forward.
The head will only move while the button is pressed.
It will stop immediately if not pressed.