IonPac CS14 Manual
Doc. No. 034848-10
Page 23 of 54
Figure 6
Isocratic Elution of Morpholine and Resolution Enhancement
Through the Addition of Solvent to the Eluent
Isocratic Eluent:
10 mM MSA / 1.2% ACN
CSRS ULTRA:
AutoSuppression
External Water Mode
Isocratic Eluent:
9 mM MSA
CSRS ULTRA:
AutoSuppression Recycle Mode
Isocratic Eluent:
8 mM MSA / 5% ACN
CSRS ULTRA:
AutoSuppression
External Water Mode
Analyte
mg/L
1.
Li
+
1.0
2.
Na
+
4.0
3.
NH
4
+
5.0
4.
K
+
10.0
5.
Morpholine
50.0
6.
Mg
2+
5.0
7.
Ca
2+
10.0
1 mg/L = 1 ppm
Analyte
mg/L
1.
Li
+
0.5
2.
Na
+
2.0
3.
NH
4
+
2.5
4.
K
+
5.0
5.
Morpholine
25.0
6.
Mg
2+
2.5
7.
Ca
2+
5.0
1 mg/L = 1 ppm
System:
2-mm
Flow Rate:
0.25 mL/min
Injection:
1.0 mL concentrated on an
IonPac CG14
Analyte
μ
g/L
1.
Li
+
0.5
2.
Na
+
2.0
3.
NH
4
+
150.0
4.
K
+
2.0
5.
Morpholine
2,000.0
6.
Mg
2+
2.0
7.
Ca
2+
10.0
1 µg/L = 1 ppb
0
5
10
Minutes
1
2
3
4
5
6
7
μ
S
0
10
5
1
2
3
4
5
6
7
0
10
20
μ
S
0
Minutes
10
5
1
2
3
4
5
6
7
0
10
20
μ
S
0
Minutes
10
5