Dionex ICS-900 Installation and Quick Start Instructions
20
Doc. 065214-02 1/12
The sequence is displayed on the page
(see
4. Under
Execute Application
, click
Start
5. The first injection in the sequence begins
running and is highlighted in the list (see
This completes the Dionex ICS-900 installation
and quick start instructions. For detailed operation
instructions, refer to the
Thermo Scientific Dionex
ICS-900 Ion Chromatography System Operator’s
Manual
(Document No. 065215), provided on the
Thermo Scientific Reference Library DVD
(P/N 053891).
Installation Troubleshooting
Problem:
When the Dionex ICS-900 power is
turned on for the first time, a Windows message
box appears asking for a USB configuration file
(cmwdmusb.inf).
Possible Cause:
The USB cable was connected to
the PC and the Dionex ICS-900 power turned on
before Chromeleon or Chromeleon Xpress was
installed.
Solution:
1. Click the
Cancel
button in the Windows
message box.
2. Turn off the Dionex ICS-900 power and
unplug the USB cable from the PC.
3. Install Chromeleon or Chromeleon Xpress
(see
).
4. Reconnect the USB cable to the PC and turn
on the Dionex ICS-900 power. Windows will
now automatically recognize the new Dionex
ICS-900.
Figure 40. Loaded sequence
Figure 41. Sequence Control panel: Starting the batch
Figure 42. Injection running