34 | Operation
Thermo Fisher Scientific CryoExtra CE8100 Series
6.
Press “ENTER”
The display will read “COM 1
Setup”. Use the “
” keys to
adjust and press “ENTER” to save.
7.
Press “SETUP”
The display will read “COM 1
Type”. Use the “
” keys to
adjust and press “ENTER” to save.
MODBUS ID
The Controller MODBUS ID also serves as the ASCII Unit ID.
This address can be adjusted in the MODBUS Menu and has
a range of 1 to 200.
Note:
Security Level 2 or higher is required to change the
MODBUS ID.
1.
Press “SETUP”
Controller will prompt for a
password. A flashing cursor on the
entry screen will make it clear
which digit is being changed. Use
the “
” keys to scroll to the appropriate number. Press
“ENTER” to advance the cursor to the next position.
2.
Press “ENTER”
The display will read “Temperature
Menus”.
3.
Press “SETUP”
Until the display reads “Advanced
Settings”.
4.
Press “ENTER”
The display will read “Advanced Fill
Menus”.
5.
Press “SETUP”
Until the display reads “MODBUS
Menu”.
6.
Press “ENTER”
The display will read “MODBUS
ID”. Use the “
” keys to
adjust and press “ENTER” to save.
One Fill All Fill (OFAF)
This section describes how to adjust the various OFAF
settings. For more information and instructions on how to
setup an OFAF network, refer to the
section of this manual.
Note:
Security Level 3 or higher is required to adjust the OFAF
settings.
1.
Press “SETUP”
Controller will prompt for a
password. A flashing cursor on the
entry screen will make it clear
which digit is being changed. Use
the “
” keys to scroll to the appropriate number. Press
“ENTER” to advance the cursor to the next position.
2.
Press “ENTER”
The display will read “Temperature
Menus”.
3.
Press “SETUP”
Until the display reads “Advanced
Settings”.
4.
Press “ENTER”
The display will read “Advanced Fill
Menus”.
5.
Press “SETUP”
Until the display reads “OFAF
Menus”.
6.
Press “ENTER”
The display will read “OFAF ID”.
Use the “
” keys to adjust
the OFAF ID and press “ENTER” to
save.
7.
Press “SETUP
”
The display will read “OFAF Units”.
Use the “
” keys to adjust
the number of units and press
“ENTER” to save.
8.
Press “SETUP”
The display will read “One Fill All
Fill”. Use the “
” keys to
adjust the type and press
“ENTER” to save.
Restore Default Settings (Global
Password)
The Controller can be reset to the factory defaults in the
Advanced Settings Menu. This will also reset post-factory
calibrations and
global password to 3 4 5 6
. The accuracy
of the level and temperature measurements should be
confirmed after resetting.
COM 1 Setup 9600 N81
Use
to
adjust
Press ENTER to save
COM 1 Type ASCII
Use
to
adjust
Press ENTER to save
Higher User Level
Required use
to
ENTER Password 0000
Press ENTER for
Temperature Menus or
press SETUP for next
menu
Press ENTER for
Advanced Settings or
press SETUP for next
menu
Press ENTER for
Advanced Fill Menus or
press SETUP for next
Press ENTER for
MODBUS menu or
press SETUP for next
MODBUS ID 200
Use
to
adjust
Press ENTER to save
Higher User Level
Required use
to
ENTER Password 0000
Press ENTER for
Temperature Menus or
press SETUP for next
menu
Press ENTER for
Advanced Settings or
press SETUP for next
menu
Press ENTER for
Advanced Fill Menus
or press SETUP for next
menu
Press ENTER for OFAF
menus or press SETUP
for next menu
OFAF ID 1
Use
to
adjust
Press ENTER to save
OFAF Units 3
Use
to
adjust
Press ENTER to save
One Fill All Fill
SEQUENTIAL
Use
to adjust
Press ENTER to save