10 | Product Information
Thermo Fisher Scientific CryoExtra CE8100 Series
* Accuracy determined over range of -200
°C
to -135
°C
. Accuracy decreases slightly as range increases.
** Accuracy determined over a range of -200
°C
to 0
°C
. Accuracy decreases slightly as range increases.
Controller Physical Connections
Temperature Probes
2-pin twist lock
Input Power
5-pin DIN
Output Power / Sensors / Battery Backup
15-pin AMP
Serial Ports
RJ-45, 4-pin RS-485
Temperature Sensor
Type
2-wire Platinum RTD (Pt-1000)
Quantity
2
Resistance
1000
@ 0°C
Sensitivity
3.85
/ °C
Temperature Measurement
Resolution
0.1°C (0.2°F)
Accuracy – Single Point Calibration *
± 1.0°C (1.8°F)
Accuracy – Two Point calibration **
± 2.0°C (3.6°F)
Range
-200°C to 70°C (-328°F to 158°F)
Level Measurement
Type
Differential Pressure Sensor
Accuracy
± 0.5 in. (13 mm) LN
2
Resolution
0.1 in. (2.5 mm)
Range
3.0 in. to 48 in. (76 mm to 1219 mm)