28 | Operation
Thermo Fisher Scientific CryoExtra CE8100 Series
appropriate number. Press “ENTER” to advance the cursor to
the next position.
2.
Press “ENTER”
The display will read “Temperature
Menus”.
3.
Press “SETUP”
Until the display reads “Liquid Level
Menus”.
4.
Press “ENTER”
The display will read “High Level
Alarm”.
5.
Press “SETUP”
Until the display reads “Advanced
Level Menus”.
6.
Press “ENTER”
The display will read Auto Fill
Control”.
7.
Press “SETUP”
The display will read “Level
Offset”. Use the “
” keys to
adjust value and press “ENTER” to
save the change.
Additional Feature Settings
This section details how to view and adjust the settings for the
Battery Backup, Power Failure Alarm and Hot Gas Bypass.
Note:
Security Level 2 or higher is required to adjust the Hot
Gas Bypass settings.
Battery Backup Status
1.
Press “SETUP”
Controller will prompt for a
password. A flashing cursor on
the entry screen will make it clear
which digit is being changed. Use
the “
” keys to scroll to the appropriate number. Press
“ENTER” to advance the cursor to the next position.
2.
Press “ENTER”
The display will read
“Temperature Menus”.
3.
Press “SETUP”
The display will read “Add-on
Menus”.
4.
Press “ENTER”
The current battery status will be
displayed. The display will read
On AC Power if battery is
connected or not connected and
Controller running on AC.
-OR-
The display will read On Battery
Backup 26 VDC or 80% or 100%
if Controller is running on battery
backup (AC disconnected).
5.
Press “SETUP”
The Power Failure Alarm status
will display. Use the “
” keys
to ENABLE or DISABLE and press
“ENTER” to save the change.
Hot Gas Bypass Settings
1.
Press “SETUP”
Controller will prompt for a
password. A flashing cursor on
the entry screen will make it clear
which digit is being changed. Use
the “
” keys to scroll to the appropriate number. Press
“ENTER” to advance the cursor to the next position.
2.
Press “ENTER”
The display will read “Temperature
Menus”.
3.
Press “SETUP”
The display will read “Add-on
Menus”.
4.
Press “ENTER”
The current battery status will be
displayed.
Press ENTER for
Temperature Menus or
press SETUP for next
menu
Press ENTER for
Liquid Level Menus or
press SETUP for next
menu
High Level Alarm 8.0 in
Use
to
adjust
Press ENTER to save
Press ENTER for
Advanced Level Menus
or press SETUP for next
menu
Auto Fill Control
ENABLED
Use
to adjust
Press ENTER to save
Level Offset 1.3 inches
Use
to adjust
Press ENTER to save
Higher User Level
Required use
to
ENTER Password 0000
Press ENTER for
Temperature Menus or
press SETUP for next
Press ENTER for Add-on
menus or press SETUP
for next menu
Battery Status
On AC Power
Battery Status On
Battery Backup 26 VDC
80% or 100%
Power Failure Alarm
Enabled
Use
to
adjust
Press ENTER to save
Higher User Level
Required use
to
ENTER Password 0000
Press ENTER for
Temperature Menus or
press SETUP for next
menu
Press ENTER for Add-
on menus or press
SETUP for next menu
Battery Status on AC
Power