7.
Place the light cube into one of the empty slots in the back of the device.
8.
Using the tool provided on the inside of the back panel (red arrow in Figure A), secure the light
cube by tightening the two screws on the end of the cube (red arrows in Figure B).
A
B
Note:
Figure B shows two light cubes installed.
9.
To remove a light cube, unscrew both screws that secure it to the instrument.
10.
Thread the light cube removal tool into the central hole in the cube (white arrows in Figure B) and
gently pull the light cube out of the device.
Note:
Always store the cube removal tool in the back panel for easy access.
11.
Install the back panel and secure it in its place with both ¼-turn fasteners.
12.
Plug the power cord back into the Countess
™
3 FL Automated Cell Counter.
13.
Turn on the Countess
™
3 FL Automated Cell Counter by flipping the power switch on the back of
the instrument to the ON position.