3.
Gently pipet 10 µL of the sample into the sample inlet, allowing capillary action to draw the sample
into the counting chamber. A properly loaded counting chamber should have a thin, even film of
fluid under the cover slip.
Note:
Each chamber in the Countess
™
Cell Counting Chamber Slide or the Countess
™
3 FL
Reusable Slide has a 10
‑
µL sample capacity. Do not overfill the slide chambers.
4.
To use the Countess
™
Reusable Slide, unlatch the back panel of the Countess
™
3 FL Automated
Cell Counter with the two captive ¼-turn fasteners that secure the back panel on the rear of the
instrument.
Chapter 4
Slide preparation and operation
Load the Countess
™
Reusable Slide
4
22
Countess
™
3 FL Automated Cell Counter User Guide