Note:
The light source buttons select the light channel (brightfield and/or fluorescence) for sample
illumination and are used when setting the exposure for the selected channel (Steps 9-11); they do
not determine which channels are used for capturing the image.
7.
To set exposure, press
Adjust
to go to the
Adjust
screen.
8.
Press the light source button for the channel you wish to set exposure and adjust the exposure
using the light source slider. Repeat the procedure for the remaining channels, if desired.
9.
Press
Apply
to keep the current setting and return to the
Setup
screen OR press
Cancel
or
Back
to return to the
Setup
screen without saving changes .
Chapter 6
Fluorescence assays
Count cell fluorescence
6
Countess
™
3 FL Automated Cell Counter User Guide
31