The plastic control panel frame is attached to the cabinet by Velcro strips.
Grasp the frame by the corners and pull it off. There are also small indents
located along the edges of the panel to accommodate a flat screwdriver
blade. The display circuit board is attached to the back of the control panel
by ten 6-32 x 3/8” Phillips head screws.
All circuit boards (with the exception of the control panel on the front of
the unit) are installed on a swing-down electronics panel inside the rear
portion of the cabinet.
To access the panel, unplug the shaker and move it to a sturdy location
that will allow the back of the cabinet to swing down and lie flat. Remove
the screws indicated by the arrows in Figure 4-2 and lower the back panel.
4-4
Orbital Shaker
Thermo Scientific
Section 4
Service
Troubleshoot/Replace
Circuit Boards
Electronics Panel
Figure 4-2.