Toggle the Sensor Fault alarm with either the up (On) arrow or the down
(Off ) arrow. Pressing the Next (temperature) button advances the display
to the next alarm, saving the on/off setting to memory.
If no buttons are pressed, the display automatically returns to the
Operating Screen after about fifteen seconds, saving the selection to
memory.
Toggle the Temperature High/Low alarm with either the up (On) arrow or
the down (Off ) arrow. Pressing the Return (temperature) button advances
the display to the display shown below.
If no buttons are pressed, the display automatically returns to the
Operating Screen after about fifteen seconds, saving the selection to
memory.
Whether the shaker has been operated in the hold or countdown modes,
and/or has been turned off and unplugged many times, the microprocessor
control system maintains a running total operating hours.
To view this information, access
the Configuration menu by
pressing the down arrow, up
arrow and Silence button in
sequence, then press the down
arrow three times.
Orbital Shaker 2-17
Thermo Scientific
Section 2
Operation
Sensor Fault
Temperature High or Low
View Total Operating
Hours