Pressing RunHrs shows the total accumulated run hours as displayed in the
illustration below. In about fifteen seconds, the display returns to the
Operating Screen.
Heat percent is intended for factory use only, but can be helpful in
troubleshooting the heat control system.
To view this information, access
the Configuration menu by
pressing the down arrow, up
arrow and Silence button in that
sequence, then press the down
arrow three times.
Press Heat %.
Heat % is the percentage of time that the heater is turned on during a five
second period. Example: If the heater is being cycled on for two seconds
and off for three seconds, the Heat % value is 40 percent.
Return to RunHrs Rmte Heat% screen by pressing any of the three
buttons beneath the display.
If no buttons are pressed, the display automatically returns to the
Operating Screen in about fifteen seconds.
2-18
Orbital Shaker
Thermo Scientific
Section 2
Operation
View Total Operating
Hours (continued)
Heat %