1-2
Orbital Shaker
Thermo Scientific
Section 1
Installation
If the shaker is to be operated at a temperature 5° to 10°C above ambient,
the circular vent at the rear of the cabinet must be 1” from the wall and
the rear vent must be fully opened by turning the Phillips screw
clockwise. Refer to Figure 1-2.
If operating temperatures are more than 10° above ambient, the rear vent
must remain closed (fully counterclockwise).
A power module (Figure 1-3) located on the right side of the shaker
contains the On/Off power switch, two fuses, and the line cord receptacle.
It is the mains disconnect for the unit.
Figure 1-4 lists the voltage, rating and part numbers of the power module
fuses.
The electrical power requirements of the MaxQ SHKE420HP shaker are:
115VAC, 50/60 Hz, 1 PH, 6.5 FLA
230VAC, 50/60 Hz, 1 PH, 3.2 FLA
Special Ventilation
Requirements
Power
Module
(mains
disconnect)
ORBITAL SHAKER
Vent
Figure 1-2.
Vents
Electrical
Requirements
Power Module
115V
230V fuse 4A P/N 230165
115V fuse 10A P/N 285633
Figure 1-3.
Power Module
Figure 1-4.
Fuses