
Using energy screens
Depending on the types of detectors you use, the system may include
a set of four energy screens. These metal screens help prevent
detector saturation and signal distortion by blocking out a portion of
the energy at all frequencies of the infrared beam. If your experiments
deal with information from a broad range of frequencies, these
screens may be the most effective means of reducing the light level.
The energy screens are labeled A, B, C and D. The following table
shows the percentage of the infrared energy that each screen passes.
It also lists the detectors typically used with each screen, as a starting
point for correcting linearity problems.
Screen %
Transmitted
*
Detectors Typically
Used With Screen
None 100
DTGS
A 30
MCT-A
B 10
C 3
D 1
*
These are nominal values that may vary due to diffraction and
detector variations.
To correct for photometric accuracy, you may need to add a
“heavier” screen (one that transmits less infrared energy). With
corrections for photometric accuracy you will notice some reduction
in the signal-to-noise ratio but will obtain more reliable quantitative
data. Generally, the signal-to-noise ratio is reduced less than is the
signal intensity.
Nicolet 380 User’s Guide
57
Содержание Nicolet 380
Страница 1: ......
Страница 6: ...iv ...
Страница 135: ...x X axis 55 XT KBr 55 y Y axis 37 69 72 84 z zoom 85 Nicolet 380 User s Guide 129 ...