Follow these steps to collect the sample spectrum:
1. Choose Open Configuration from the File menu.
The Open Configuration dialog box appears listing the available
configuration files. Opening one of these files lets you quickly
set a number of software options.
2. Select the DEFAULT.CON file and then choose OK.
This file was provided with the OMNIC software for setting the
options for this procedure.
3. If it is not already selected, select the Default - Transmission
experiment (or the Transmission E.S.P. experiment).
First click the arrow button at the right end of the Experiment
drop-down list box to display a list of available experiments.
The “Default - Transmission” experiment appears in the list.
Select the experiment by clicking its name in the list.
The parameters in the Experiment Setup dialog box are now set
correctly for collecting a polystyrene sample spectrum. These
parameters determine how OMNIC collects background and
sample spectra. For example, in the experiment you selected, the
Collect Background Before Every Sample option is selected.
This option prompts you to collect a background before
collecting the sample spectrum.
Nicolet 380 User’s Guide
23