Checking
the desiccant
If the spectrometer is sealed and desiccated, a bag of desiccant.
accessible through a hatch on the top of the instrument, protects the
beamsplitter and other optical components by reducing the amount
of water vapor inside spectrometer.
If the spectrometer is in storage
, check the round, blue indicator on
the desiccant holder under the beamsplitter compartment cover
monthly. Change the desiccant if the indicator has turned pink. Keep
all covers closed tightly while the spectrometer is in storage.
See “Spectrometer Help Topics if you need to change the desiccant.
If the spectrometer is in use
, check the humidity level inside it
monthly (more often in very humid climates) by clicking the Check
Desiccant button on the Diagnostic tab of the Experiment Setup
dialog box. If the level exceeds the factory-specified limit, a message
informs you. Be sure to change the desiccant so that the optics will
be protected.
If the spectrometer is in storage
, check the round, blue indicator on
the desiccant holder under the beamsplitter compartment cover
monthly. Change the desiccant if the indicator has turned pink. Keep
all covers closed tightly while the spectrometer is in storage.
Note
The Check Desiccant button measures humidity more accurately
than the indicator. For this reason you should rely on the button to
monitor the performance of the desiccant when the spectrometer is in
use. Rely on the indicator only when the spectrometer is in
storage.
S
See “Ordering parts” in Spectrometer Help Topics to find the part
number for ordering desiccant bags.