A spectrum is too noisy,
continued…
The detector is saturated.
Add a bandwidth-limiting filter
or an energy screen to reduce the
signal strength.
The resolution was too high.
Reduce the resolution (use a
higher numerical setting) on the
Collect tab of the Experiment
Setup dialog box.
The moving mirror velocity is set too
high for the detector.
Set Velocity on the Bench tab of
the Experiment Setup dialog box
to a lower value.
The sample in an ATR experiment is
not in good contact with the ATR
crystal.
Reapply the sample to obtain
more consistent contact between
the sample and the ATR crystal.
Nicolet 380 User’s Guide 121