Problem Possible
Cause
Solution
There is frost on the outside of
the detector.
The insulating vacuum chamber of
the dewar has begun to leak.
If you suspect the detector has a
vacuum leak, view “Maintaining
detector dewars” in the
“Maintaining your spectrometer”
book in Spectrometer Help
Topics.
A message says that data cannot
be collected.
The spectrometer is turned off.
Turn on the spectrometer and
wait until the start-up
diagnostics finish before
restarting OMNIC.
The data cable is loose, disconnected
or damaged.
Shut down the computer and
turn off the spectrometer. Check
the data cable. If it is damaged,
replace it. If the cable is loose or
disconnected, reconnect it. Turn
on the spectrometer and wait
until the spectrometer start-up
diagnostics finish. Then turn on
the computer and restart
OMNIC.
The computer port used for
communicating with the
spectrometer is not operating
correctly.
Make sure that the computer
USB 2.0 port is operating
correctly. Use the instructions
that came with the computer.
A component has failed.
Contact Thermo Electron for
service.
continued…
Thermo Electron Corporation
108