122
Results
When Go-NoGo is running, the pass/fail ratio is
displayed in the bottom right-hand corner. The first
digit represents the number of failed tests, and the
right hand digit represents the number of tests.
Ratio:
fail / test
Maximum
position
Minimum
position
Tolerance
Exit the
Application
To exit the application, press
Break
.
Using the Go-
NoGo Output
To output the Go-NoGo results to an
external device, the Go-NoGo rear
panel terminal (open collector) can be
used. The Go-NoGo terminal will
output a positive pulse each time a
NoGo violation has occurred for a
minimum of 500us. The voltage of the
pulse depends on the external pull-up
voltage.
Timing Diagram
Enable
NoGo
500
us
Waveform
Circuit Diagram
Go-NoGo signal
220
Ω
Go-NoGo
BNC
B
+
CMOS
(Max 16V)
Or TTL
(Max 5V)