107
4-6-11. Using the Timeout Trigger
Panel Operation 1. Press the trigger
Menu
key.
Menu
2. Press the
Type
key from the lower
bezel menu.
3. Select
Others
→
Timeout
from the side menu. The
Timeout indicator appears
at the bottom of the
display.
From left: Source, Trigger type, threshold level,
coupling
4. Press
Source
from the lower menu.
Use the side menu to select a source.
Range
Channel 1 ~ 4, D0~D15, EXT (Ext
Probe: Volt/Current, Attenuation:
1mX~1kX ), AC Line
5. Press
Trigger
When
from the lower
menu.
Then use the side menu to select trigger
conditions.
Condition Stays High, Stays Low, Either
6. Press
Level
from the lower bezel to
set the trigger level.
Range
-XXV~XXV
Set to TTL 1.4V
Set to ECL -1.3V
Set to 50%
7. Press
Timer
from the lower bezel
to set the timer time.
Range
-10ns~10.0S