95
3. Press
Attenuation
on the side
menu and use the Variable knob to
set the attenuation.
Alternatively, press
Set to 10X
.
Range
1mX ~1kX (1-2-5 step)
Note
The attenuation factor adds no influence on the
real signal. It just changes the voltage/current
scale on the display.
4-5-10. Set the Deskew
Background
The deskew function is used to compensate for the
propagation delay between the oscilloscope and
the probe.
Panel Operation 1. Press one of the
Channel
keys.
CH1
2. Press
Probe
from the bottom
menu.
3. Press
Deskew
on the side menu
and use the Variable knob to set
the deskew time.
Alternatively, press
Set to 0s
to
reset the deskew time.
Range
-50ns~50ns, 10ps increments
4. Repeat the procedure for another channel if
necessary.
4-6. Trigger
The trigger configures the conditions for when the DCS-9700
captures a waveform.
The following trigger overview only applies to the analog channels,
for triggering details using the optional logic analyzer module,
please see the DCS-9700 Options User Manual for details.
4-6-1. Trigger Type Overview
Edge
The edge trigger is the simplest trigger type. An
edge trigger triggers when the signal crosses an
amplitude threshold with either a positive or
negative slope.
Rising edge trigger