121
2. Press
Go Back
to return to the
previous menu.
Set Boundary
Tolerance
1. To set the Go-NoGo boundary
tolerance, press
Reference Mode
.
Auto Tolerance 2. To set the boundary tolerance as a
percentage offset from the source
waveform, press
Auto Tolerance
and
use
the Variable knob.
VARIABLE
Offset
0.4% ~ 40% (.4% steps)
Maximum and
Minimum
Position
3. To manually set the template
tolerance, press
Minimum Position
or
Maximum Position
and use the
Variable knob to set the absolute
minimum or maximum position.
or
Range
Voltage division range
Save Boundary
Template
4. Press
Save Operation
to save the
tolerance boundaries.
5. The Maximum Position tolerance
will be saved to reference
waveform R1, and the Minimum
Position tolerance to R2.
6. Press
Go Back
to return to the
previous menu.
Note
It may be necessary to save a reference waveform
to R1 and R2 before the tolerance for Maximum
and Minimum Position can be created. This can be
done by first creating and saving the Auto
Tolerance boundary.
Start Go-NoGo Press
Enable
to start the Go-NoGo
test. The Enable button will change to
Disable. Pressing
Disable
will stop the
Go-NoGo test and toggle the button
back to Enable.
If the Violating setting was set to Stop
or Stop Beep, press
Enable
to restart
the test after it has stopped.