30
3-1-2. Using Autoset
Background
The Autoset function automatically configures the
panel settings to the best viewing conditions, in the
following way.
Selecting the horizontal scale
Positioning the waveform horizontally
Selecting the vertical scale
Positioning the waveform vertically
Selecting the trigger source channel
Activating the channels
Procedure
1. Connect the input signal to
the oscilloscope and press
the Autoset key.
Autoset
2. The waveform(s) appears in the center of the
display.
Undo option
Before Autoset
After Autoset
Undoing the
Autoset
To undo the Autoset, press
Undo
(available for a few
seconds).
Undo
Adjusting the
trigger level
If the waveform is still unstable,
try adjusting the trigger level up
or down by using the Trigger
Level knob.
LEVEL
Limitation
Autoset does not work in the following situation.
Input signal frequency less than 30Hz
Input signal amplitude less than 30mV
Содержание DCS-4605
Страница 1: ...INSTRUCTION MANUAL DIGITAL STORAGE OSCILLOSCOPE DCS 4605 B71 0392 01 ...
Страница 12: ......
Страница 81: ...69 ...
Страница 116: ......