54
4. CONFIGURATION
The Configuration chapter describes how to configure panel settings to
make measurements and observations suited to the application needs.
4-1.
Acquisition
The acquisition process samples the analog input signals and converts
them into digital format for internal processing. You may select the
normal, average, or peak detect acquisition mode.
4-1-1. Selecting the acquisition mode
Procedure
1. Press the Acquire key.
Acquire
2. Select the acquisition mode
between
Normal
,
Average
and
Peak Detect
.
Normal
Peak
Detect
Average
Range
Normal
All of the acquired data is used to draw
the waveform.
Average
Multiple data is averaged to form a
waveform. This mode is useful for
drawing a noise-free waveform. To
select the number, press
Average
repeatedly.
Average number: 2, 4, 8, 16, 32, 64,
128, 256
Peak
detect
To activate the Peak detect mode,
press
Peak-Detect
. Only the minimum
and maximum value pairs for each
acquisition interval (bucket) are used.
This mode is useful for catching
abnormal glitches in a signal.