98
7-2.
DCS-4605 Specifications
The specifications apply when the oscilloscope is powered on for at
least 30 minutes under +20°C~+30°C.
7-2-1. Specifications
Vertical
Sensitivity
2mV/div~10V/div (1-2-5 increments)
Accuracy
± 3% Full scale
Bandwidth
(
–3dB)
DC coupling: DC ~ 50MHz
AC coupling: 10Hz ~ 50MHz
Rise Time
< 7ns approx.
Input Coupling
AC, DC, Ground
Input Impedance
1MΩ±2%, ~15pF
Polarity
Normal, Invert
Maximum Input
300V rms, CAT II
Math Operation
+,
–, FFT
Offset Range
2mV/div
100mV/div
1V/div
10V/div
~ 50mV/div
~ 500mV/div
~ 5V/div
: ±0.4V
: ±4V
: ±40V
: ±300V
Bandwidth Limit
20MHz (−3dB)
Trigger
Sources
CH1, CH2, Line, External
Modes
Auto / Normal / Single , Edge / TV /
Pulse
Coupling
AC, DC, LF rej, HF rej, Noise rej
Trigger Sensitivity 0.5div or 5mV (DC ~ 25MHz)
1.5div or 15mV (25MHz~50MHz)
External trigger
Range
DC: ±15V, AC: ±2V
External Trigger
Sensitivity
~ 50mV (DC~25MHz)
~ 100mV (25MHz~50MHz)
Input Impedance
1MΩ±2%, ~15pF
Maximum Input
300V rms, CATII
Horizontal
Range
1ns/div~50s/div, 1-2.5-5 increment
Roll: 50ms/div
– 50s/div
Modes
Main, Window, Window Zoom, Roll, X-Y
Accuracy
±0.01%
Pre-Trigger
10 div maximum
Post-Trigger
1000 div
X-Y Mode
X-Axis Input
Channel 1
Y-Axis Input
Channel 2
Phase Shift
±3° at 100kHz
Signal Acquisition Real-Time
250MS/s maximum
Equivalent
25G S/s maximum
Vertical
Resolution
8 bits