44
3-5.
Go No-Go Testing
3-5-1. Overview
Background
Go-NoGo testing checks if a waveform conforms to
a user-specified maximum and minimum boundary
(template). The testing can be set to stop or
continue each time the template has or has not
been violated by the input waveform.
Settings
Item
Default
Details
NoGo criteria: When inside
or outside the boundary
Inside
Page 44
Source
Channel 1 Page 45
Test continue or stop when
NoGo occurs
Stop
Page 45
Boundary (template)
–
selects the minimum and
maximum boundaries
(template) from a single
waveform
Auto
(0.4%)
Page 46
Run Tests
Page 50
3-5-2. Edit: NoGo When
Procedure
1. Press the Utility key.
Utility
2. Press the
More
key.
More
3. Press
No Go When
repeatedly to select the
NoGo conditions.
No Go When
NoGo when the waveform is
inside the boundary (template)
NoGo when the waveform is
outside of the boundary
(template)