FAQ
• The input signal does not appear in the display.
• I want to remove some contents from the display.
• The waveform does not update (frozen).
• The probe waveform is distorted.
• Autoset does not catch the signal well.
• I want to clean up the cluttered panel settings.
• The accuracy does not match the specifications.
The input signal does not appear in the display.
Make sure you have activated the channel by pressing the CH key
(page 29).
I want to remove some contents from the display.
To clear the math result, press the Math key again (page 41).
To clear the cursor, press the Cursor key again (page 39).
To clear the Help contents, press the Help key again (page 28).
The waveform does not update (frozen).
Press the Run/Stop key to unfreeze the waveform. See page 31 for
details. For trigger setting details, see page 65.
If this does not help, press the CH key. If the signal still does not
appear, press the Autoset key.
The probe waveform is distorted.
You might need to compensate the probe. For details, see page 95.
Note that the frequency accuracy and duty factor are not specified
for probe compensation waveforms and therefore it should not be
used for other reference purposes.
Autoset does not catch the signal well.
The Autoset function does not catch signals well under 30mV or
30Hz. Please operate the oscilloscope manually.