82
3-14-9. :TRIGger:HLEVel
Set
Query
Description
Sets or queries the high trigger level.
Note
Applicable for Rise and Fall/Pulse Runt triggers.
Syntax
:TRIGger:HLEVel {TTL | ECL| | <NRf> | ?}
Related commands :TRIGger:TYPe
Parameter
<NRf>
High level value.
TTL
Sets the high trigger level to TTL.
ECL
Sets the high trigger level to ECL.
Return parameter
<NR3>
Returns the trigger high level.
Example1
:TRIGger:HLEVel TTL
Sets the trigger high level to TTL.
Example2
:TRIGger:HLEVel 3.30E-1
Sets the trigger high level to 330mV/mA.
3-14-10. :TRIGger:LLEVel
Set
Query
Description
Sets or queries the low trigger level.
Note
Applicable for Rise and Fall/Pulse Runt triggers.
Syntax
:TRIGger:LLEVel {TTL | ECL| <NRf> | ?}
Related commands :TRIGger:TYPe
Parameter
<NRf>
Low level value.
TTL
Sets the low trigger level to TTL.
ECL
Sets the log trigger level to ECL.
Return parameter
<NR3>
Returns the trigger low level.
Example1
:TRIGger:LLEVel TTL
Sets the trigger low level to TTL.
Example2
:TRIGger:LLEVel -3.30E-3
Sets the trigger low level to -330mV/mA.