22
3-4. Vertical Commands
3-4-1. :CHANnel<X>:BWLimit ............................................................................ 22
3-4-2. :CHANnel<X>:COUPling .......................................................................... 22
3-4-3. :CHANnel<X>:DESKew ........................................................................... 23
3-4-4. :CHANnel<X>:DISPlay ............................................................................. 23
3-4-5. :CHANnel<X>:EXPand ............................................................................ 23
3-4-6. :CHANnel<X>:IMPedance? ...................................................................... 24
3-4-7. :CHANnel<X>:INVert ............................................................................... 24
3-4-8. :CHANnel<X>:POSition ........................................................................... 24
3-4-9. :CHANnel<X>:PROBe:RATio ................................................................... 25
3-4-10. :CHANnel<X>:PROBe:TYPe .................................................................. 25
3-4-11. :CHANnel<X>:SCALe ............................................................................ 25
3-4-1. :CHANnel<X>:BWLimit
Set
Query
Description
Sets or returns the bandwidth limit on/off.
Syntax
:CHANnel<X>:BWLimit {FULL | <NR3> | ?}
Parameter
<X>
Channel 1,2,3,4
FULL
Full bandwidth
<NR3>
Sets the bandwidth limit to a pre-defined
bandwidth.
100E+6: 100MHz
20E+6: 20MHz
Return Parameter <NR3>
Returns the bandwidth.
Full
Full bandwidth
Example
:CHANnel1:BWLimit 2.000E+07
Sets the channel 1 bandwidth 20MHz
3-4-2. :CHANnel<X>:COUPling
Set
Query
Description
Selects or returns the coupling mode.
Syntax
CHANnel<X>:COUPling {AC | DC | GND | ?}
Parameter
<X>
Channel 1,2,3,4
AC
AC coupling
DC
DC coupling
GND
Ground coupling
Return parameter Returns the coupling mode.
Example
:CHANnel1:COUPling DC
Sets the coupling to DC for Channel 1.